Application of Optical NDE in Design Verification to Assess the Elastic Recoil and Foreshortening of Coronary Stents
Published 22-09-2024
Keywords
- Optical,
- NDE,
- Biocompatible implant,
- Stent,
- Quality assurance
How to Cite
Copyright (c) 2024 Journal of Non-Destructive Testing & Evaluation (JNDE)

This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.
Abstract
Coronary stenting has emerged has the most favoured approach for treating Coronary Artery Disease. This study examined optical non-destructive evaluation methods to verify the critical dimensions of coronary stents, which are essential in helping clinicians select appropriate stents for individual patients. The focus was on assessing the elastic recoil and foreshortening of Chitra Coronary stents. The researchers conducted dimensional verification of the Chitra Coronary stent in its unmounted, mounted, and deployed states using a digital microscope and in situ laser micrometer measurements during stent-balloon expansion characterization studies as per international standards. The findings indicated that the recoil and foreshortening of the Chitra Stent met the product acceptance criteria. The study also underscored the pivotal significance of optical non-destructive assessment in guaranteeing the safety characteristics of stents during the design stage.
References
- H. Arora, A. Gupta, and D. Gupta, “Impact of Coronary Stent Price Regulation Section on the Interventional Cardiology Landscape of India: An Assessment of Health Economics Outcomes,” J. Clin. Diagnostic Res., vol. 16, no. 3, pp. 3–6, 2022, doi: 10.7860/jcdr/2022/55038.16116.
- N. W. Bressloff, G. Ragkousis, and N. Curzen, “Design Optimisation of Coronary Artery Stent Systems,” Ann. Biomed. Eng., vol. 44, no. 2, pp. 357–367, Feb. 2016, doi: 10.1007/s10439-015-1373-9.
- M. Y. Ho et al., “The development of coronary artery stents: From bare-metal to bio-resorbable types,” Metals, vol. 6, no. 7. MDPI AG, Jul. 01, 2016, doi: 10.3390/met6070168.
- T. Simard, B. Hibbert, F. D. Ramirez, M. Froeschl, Y. X. Chen, and E. R. O’Brien, “The Evolution of Coronary Stents: A Brief Review,” Canadian Journal of Cardiology, vol. 30, no. 1. pp. 35–45, Jan. 2014, doi: 10.1016/j.cjca.2013.09.012.
- S. Garg and P. W. Serruys, “Coronary stents: Current status,” Journal of the American College of Cardiology, vol. 56, no. 10 SUPPL. Aug. 31, 2010, doi: 10.1016/j.jacc.2010.06.007.
- D. Stoeckel, C. Bonsignore, S. Duda, D. Stoeckel, C. Bonsignore, and S. Duda, “A survey of stent designs ,” Minimally Invasive Therapy & Allied Technologies, 11(4), 137–147,2002..
- F. Migliavacca, L. Petrini, M. Colombo, F. Auricchio, and R. Pietrabissa, “Mechanical behavior of coronary stents investigated through the finite element method,” 2002.
- H. Li et al., “Multi-objective optimization of coronary stent using Kriging surrogate model,” Biomed. Eng. Online, vol. 15, Dec. 2016, doi: 10.1186/s12938-016-0268-9.
- Sujesh, S., et al. "An Automated Instrumentation System for Expansion and Deployment Characterization of Tubular Stent Devices." International Conference on System Dynamics and Control-ICSDC. Vol. 19. No. 22. 2010.
- PMA P210014: FDA Summary of Safety and Effectiveness Data, https://www.accessdata.fda.gov/cdrh_docs/pdf21/P210014B.pdf , Accessed on 30/07/2024.
- ASTM International. F2081-06, Standard Guide for Characterization and Presentation of the Dimensional Attributes of Vascular Stents, 2017, ASTM International, West Conshohocken, PA, USA, https://doi.org/10.1520/F2081-06R17.